No.8

Introduction
Data Policy
Data
Documentation
Contact Information



Zugspitze

-CO-

back
Station Contact Person(s) Observation Data and Quick Plot

Observation
Category : Air sampling observation
Situation : 31.03.2002
Time zone : Local time 1


Sampling
Sampling height :
Description : continuous
Sampling and analysis frequency : Continuous
Sampling environment :
Description for sampling analysis : Air inlet


Instrument and Analysis
Measurement method : NDIR
Current status and history of instrument : NDIR
Description of instrument : Thermo Environmental Instruments Inc.
TE 48S


Calibration
Current scale employed in the measurement : NOAA/CMDL Scale
Measurement calibration : Calibration gases: CO in natural air
Scale and calibration(treasability) : NOAA/CMDL Scale


Data Processing
Measurement unit : mg/m3-25C
Data processing : All values are quality controlled
Processing for averaging : Processing for Hourly Data:
Only technically biased values are removed (caused by calibration or malfunctioning of equipment).
Processing for Daily Data:
Take the arithmetic mean of selected harfhourly data if at least 2/3 of the halfhourly values of one day are available.
Processing for Monthly Data:
Take the arithmetic mean of selected daily data if at least 2/3 of the daily values of one month are available.
Data flag : Flags put on hourly data are as follows.
1: valid value
6: calibration
9: error code
Data remarks :


Other Information
Scientific aim :
Reference : "Qualitatssicherungs-Handbuch des UBA-Messnetzes", UBA Texte 28/04, (Handbook for quality assurance for the measurement network for the Federal Environmental Agency, Germany), Dr.Klaus Nienerowski, Umweltbundesamt, Berlin, Juni 2004

http://www.umweltbundesamt.de/luft/luftmessnetze/ubamessnetz.htm


submitted by Federal Environmental Agency Germany