Hok Tsui
-CO-
Observation |
Category |
: |
Air sampling observation |
Situation |
: |
|
Time zone |
: |
Local time +8 |
Sampling |
Sampling height |
: |
|
Description |
: |
continuous |
Sampling and analysis frequency |
: |
Hourly |
Sampling environment |
: |
|
Description for sampling analysis |
: |
Real time point analyzer. The sampling line is 1/4" teflon tube. The sampling inlet is mounted on a scaffolding placed at 2m above roof top. |
Instrument and Analysis |
Measurement method |
: |
|
Current status and history of instrument |
: |
NDIR Gas filter correlation |
Description of instrument |
: |
TECO
Teco 48S |
Calibration |
Current scale employed in the measurement |
: |
|
Measurement calibration |
: |
Zeroed once everyhour using pd catalyst. Span checked once every 4 hours using dynamic dilution of Scott Marrin standard gas. Span gas concentration vary from 500 ppb to 1000 ppb. |
Scale and calibration(treasability) |
: |
0 to 2000 ppb |
Data Processing |
Measurement unit |
: |
ppb |
Data processing |
: |
|
Processing for averaging |
: |
|
Data flag |
: |
|
Data remarks |
: |
|
Other Information |
Scientific aim |
: |
|
Reference |
: |
|
|
submitted by The Hong Kong Polytechnic University |
|
|