No.8

Introduction
Data Policy
Data
Documentation
Contact Information



South Pole

-O3-

back
Station Contact Person(s) Observation Data and Quick Plot

Observation
Category : Air sampling observation
Situation : ongoing
Time zone : UTC


Sampling
Sampling height : 6
Description : continuous
Sampling and analysis frequency : Continuous
Sampling environment : clean air facility
Description for sampling analysis : teflon inlet with PTFE filter membrane
1-3 lpm sample flow rate


Instrument and Analysis
Measurement method : Light absorption analysis (UV)
Current status and history of instrument : 1975.01-1975.12 : ECC ( Electrochemical concentration cell ozone meter )
1976.01-2004.07 : DASIBI (UV- Photometer)
2004.08-: TEI 49C (UV- Photometer)
Description of instrument : 1.0 ppb precision
+/- 1.0% full scale linearity


Calibration
Current scale employed in the measurement : the NIST standard
Measurement calibration : 0-200 ppb calibration scale
sample gas produced from NIST standard
Scale and calibration(treasability) : Calibration standard February 2012


Data Processing
Measurement unit : ppb
Data processing : 1 hour records with instrument details
5 minute data average file
All data is scrutinized to only remove data that has been directly influenced by an outside affect (ie. power outages, zero span checks...)
Processing for averaging : Processing for Hourly Data:
all data
Processing for Daily Data:
all data
Processing for Monthly Data:
all data
Data flag :
Data remarks :


Other Information
Scientific aim : Background surface ozone
Antarctic depletion
Summer ozone variability
Reference : Oltmans, SJ and Levy II, H, Surface ozone measurements from a global network,
Atmos. Environ., 28, 9-24, 1994.

http://www.esrl.noaa.gov/gmd/ozwv/surfoz/index.html


submitted by Earth System Research Laboratory, NOAA



The WDCGG is operated by the Japan Meteorological Agency
in collaboration with the World Meteorological Organization