Rigi
-VOCs-
Observation |
Category |
: |
Air sampling observation |
Situation |
: |
ongoing |
Time zone |
: |
Local time +1 |
Sampling |
Sampling height |
: |
5 |
Description |
: |
continuous |
Sampling and analysis frequency |
: |
2-hour |
Sampling environment |
: |
rural environment |
Description for sampling analysis |
: |
Air Intake: open-face, ambient |
Instrument and Analysis |
Measurement method |
: |
Gas Chromatography (FID) |
Current status and history of instrument |
: |
Measurements are performed since
2001 and are on-going.
The instrument was a
Gas chromatograph:
Varian 3600 (2001-08.2005)
Agilent (08.2005-ongoing)
The chromatographic column is a PLOT (KCl) column (50m x 0.53 mm)
2 L of air are trapped at -44 °C using an ADS system, as described by Simmonds et al. (Anal. Chem., 1995) |
Description of instrument |
: |
|
Calibration |
Current scale employed in the measurement |
: |
NPL-30 VOC component mixture in the range of ~4 ppb |
Measurement calibration |
: |
manually every 4 weeks |
Scale and calibration(treasability) |
: |
the system was audited in 2005 by the WCC-VOC |
Data Processing |
Measurement unit |
: |
ppb |
Data processing |
: |
Quality assurance procedures involve time series plots, consistency checks. |
Processing for averaging |
: |
Processing for Daily Data: Estimated standard uncertainty of single injections approx. 10 %; detection limit approx. 10 ppt
|
Data flag |
: |
|
Data remarks |
: |
|
Other Information |
Scientific aim |
: |
|
Reference |
: |
|
|
submitted by Swiss Federal Laboratories for Materials Testing and Research |
|
|